Operation
Selecting the signals and ranges for
Analog1 and Analog2
Press the key »Param« repeatedly until »Analog1« is
displayed light on dark background.
Select the desired signal with the keys
Press the key »Param«, and then select
the desired scale value with the keys
Proceed in the same way for Analog2.
The following signals and ranges are available:
Airway pressure
Airway pressure
Mean airway pressure
Mean airway pressure
FiO
2
Flow
Flow
Flow
Flow
Volume
Volume
Volume
Volume
Tidal Volume
Tidal Volume
Tidal Volume
Tidal Volume
Minute Volume MV
Minute Volume MV
Minute Volume MV
Minute Volume MV
Leakage
Spont. fraction of MV
Default settings:
Analog1:
Flow
Analog2:
Airway pressure
To select these settings during operation:
Press the key »* * * * /P«.
If a measured value exceeds the the scale limits, the voltage is
limited to the end-of-scale value.
26
or
or
–10 . . . 90 mbar
–> 0 . . .10 V
–5 . . . 45 mbar
–> 0 . . .10 V
–10 . . . 90 mbar
–> 0 . . .10 V
–5 . . . 45 mbar
–> 0 . . .10 V
0 . . . 100 Vol.%
–> 0 . . .10 V
–40 . . . 40 L/min
–> 0 . . .10 V
–20 . . . 20 L/min
–> 0 . . .10 V
–10 . . . 10 L/min
–> 0 . . .10 V
–5 . . . 5 L/min
–> 0 . . .10 V
0 . . . 500 mL
–> 0 . . .10 V
0 . . . 100 mL
–> 0 . . .10 V
0 . . . 50 mL
–> 0 . . .10 V
0 . . . 25 mL
–> 0 . . .10 V
0 . . . 500 mL
–> 0 . . .10 V
0 . . . 100 mL
–> 0 . . .10 V
0 . . . 50 mL
–> 0 . . .10 V
0 . . . 25 mL
–> 0 . . .10 V
0 . . . 10 L/min
–> 0 . . .10 V
0 . . . 5 L/min
–> 0 . . .10 V
0 . . . 1 L/min
–> 0 . . .10 V
0 . . . 0,5 L/min
–> 0 . . .10 V
0 . . . 100 %
–> 0 . . .10 V
0 . . . 100 %
–> 0 . . .10 V
–20 . . . 20 L/min
–10 . . . 90 mbar
.
.