interference (which can be determined by
unplugging the power cord from the unit), try these
measures: Relocate the unit with respect to the
susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
Important EMC notices for use in the medical environments
• The UP-D55 needs special precautions regarding EMC
and needs to be installed and put into service according
to the EMC information provided in this instructions for
use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the UP-D55.
Guidance and manufacturer's declaration-electromagnetic emissions
The UP-D55 is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-D55 should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Warning
If the UP-D55 should be used adjacent to or stacked with
other equipment, it should be observed to verify normal
operation in the configuration in which it will be used.
Consult your dealer. (According to standard EN60601-
1-2 and CISPR11, Class B, Group 1)
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the UP-D55.
Compliance
The UP-D55 uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The UP-D55 is suitable for use in all establishments,
Class B
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Class A
Complies
Electromagnetic environment-guidance
GB
3