5.2 Maximum permissible load
All weighing platforms are equipped with an overload protection. However, if the load
exceeds the maximum permissible load, damage of mechanical parts may occur.
The static load-bearing capacity, i.e. the maximum permissible load, is dependent on the
type of loading (positions A – D).
Position
A central load
B side load
C one-sided corner load
individual wheel load
D
min. wheel distance 500 mm
5.3 Ambient conditions
The weighing platform and the gas pressure springs of the raisable weighing platforms
may only be operated in the range of –10 °C to + 40 °C / 14° F to 104 °F.
5.4 Digital weighing interface specification (safe area only)
Interface type
Interface protocol
Max. cable length
5.5 Connectivity to weighing terminals
Weighing interface
SICSpro
ACC409xx SICSpro-
IDNet adapter (option)
5.6 Specifications for Category 2 / Category 3
You will find the specifications for Category 2 / Category 3 in the corresponding type
approval documents, see table in section 2.3.
PUA57..CS
PUA57..FL
800 kg
400 kg
300 kg
200 kg
RS422
SICSpro
20 m
Weighing terminal
Any METTLER TOLEDO weighing terminal with SICSpro RS422
interface may be connected.
Only the following METTLER TOLEDO legacy IDNet weighing
terminals may be connected:
IND570, IND690, IND780, IND890, ID7, ICS__9, ICS__5.
PUA57..E
PUA57..FM
PUA57..FH
1850 kg
3000 kg
900 kg
1800 kg
450 kg
900 kg
400 kg
800 kg
Technical data and operating limits