Installation

Microcentrifuge 24
4
Installation
4.1

Selecting the location

WARNING! Danger due to incorrect voltage supply.
 Only connect the device to voltage sources which correspond with the electrical requirements on the name plate.
 Only use earth/grounded sockets with a protective earth (PE) conductor.
 Only use the mains/power cord supplied.
NOTICE! If an error occurs, any objects in the immediate proximity of the device may become damaged.
 In accordance with the recommendations of EN 61010-2-020, leave a safety clearance of 30 cm (11.8 in) around the device during
operation.
 Please remove all materials and objects from this area.
NOTICE! Damage due to overheating.
 Do not install the device near heat sources (e.g. heating, drying cabinet).
 Do not expose the device to direct sunlight.
 Ensure unobstructed air circulation. Maintain a clearance of at least 30 cm (11.8 in) around all ventilation gaps.
NOTICE! Radio interference.
For devices with Class A noise emission in accordance with EN 61326-1/EN 55011, the following applies: This devices has been developed
and tested in accordance with CISPR 11 Class A. The device may cause radio interference in domestic environments and is not intended for
use in residential areas. The device cannot ensure adequate protection of radio reception in residential areas and domestic environments.
 If necessary, take appropriate measure to eliminate the interferences.
Mains/power connection for centrifuges: The operation of the centrifuge is only permitted in a building installation which complies with
the applicable national regulations and standards. In particular, it needs to be ensured that there are no prohibited loads on the supply
lines and assemblies that are located before the internal protection of the device. This can be ensured by additional circuit breakers or
other suitable fuse elements in the building installation.
The mains/power switch and the disconnecting device of the mains/power line must be easily accessible during operation
(e.g. a residual current circuit breaker).
Select the location of the device according to the following criteria:
• Mains/power connection in accordance with the name plate.
• Minimum distance to other devices and walls: 30 cm (11.8 in).
• Resonance free table with horizontal even work surface.
• The surrounding area must be well ventilated.
• The location is protected against direct sunlight.
Do not use this device near strong electromagnetic sources (e.g. unshielded high frequency sources) as they could impede proper
functioning of the device.
14
· Microcentrifuge 24 · Manual