Problem solving: If any of the tests described above failed, you
should change the connecting cable and repeat the test. If the test
still shows the same result, the transmitter or the cable is defective.
If you are using a transmitter with current outputs, so ensure the out-
puts are correctly wired.
You must never do a calibration (the simulators are read-only).
You must never increment autoclave cycles (always set "NO"
to cycle counter), this is only needed if autoclave has been set to
a value higher than 0.
Different transmitters may display different values and errors.
Out of range behavior is also different depending on the transmitter
in use thus different values/readings may occur.
Set the transmitter input filter to off (only for SIM O
ppb TGL and SIM O
Do not use the O
on the label of each plug-in head. Please send them back to
METTLER TOLEDO for recertification.