The displayed weight is
permanently changing
The weighing result is
obviously incorrect

14.1 Error messages

E
Weighing range exceeded; the set load exceeds the weighing range.
Unload the balance.
E
Weighing range exceeded, e.g. weighing plate not installed.
CAL-no
Adjustment error.
32
• Draught/air movement
• Table/floor vibrations
• Weighing pan has contact with other objects.
• Electromagnetic fields / static charging (choose a
different location / switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• The balance is on an uneven surface.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose a
different location / switch off interfering device if
possible)
ADB_ADJ-BA-e-1830