Rack controller
Vorschau ausblenden Andere Handbücher für Turbo-V 301-AG:
Inhaltsverzeichnis

Werbung

The following parameters can be changed:
Tab. 9
BYTES
1
2-5
6
7
8
9
When the Request is "I" (Status readings), the Answer will contain a
string of 2 characters with the following parameters:
Tab. 10
BYTES
2
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
1
MSB
LSB
- - X X X X X X
Letter Protocol Description
MEANING
Speed threshold
Run up time in seconds (long coded in 4 bytes)
Deat time (0 = NO 1 = YES)
Reserved
Soft Start mode (0 = YES 1 = NO)
CRC
MEANING
0
= STOP
1
= WAITING INTERLOCK
2
= STARTING
3
= NORMAL OPERATION
4, 5 = HIGH LOAD
6
= FAILURE
7 = APPROACHING LOW SPEED
R2 status 0 = OFF
1 = ON
R1 status 0 = OFF
1 = ON
CRC
(F)
Technical Information
233/294
17