E. Sterilization
Moist heat/steam sterilization is the recommended method for the Symmetry Sharp Kerrison
Reusable Tips for the Symmetry Sharp Kerrison
The Symmetry Sharp Kerrison
Always consult and follow the sterilizer manufacturer's instructions for equipment operation. Sterilizing
equipment should have demonstrated efficacy (e.g. FDA approval, process validation, maintenance records).
Individual devices may be packaged in a medical grade sterilization pouch, wrap, or sterilization containers.
Care should be used when packaging so that the pouch or wrap is not torn. Devices should be wrapped
using the AAMI double wrap or equivalent method. The wrap should be cleared for use in the 132°C, 4
minutes pre-vacuum steam sterilization cycle.
Follow the load configurations specified by the sterilization container and sterilizer manufacturers. The
handle and reusable tips may be sterilized with a combined load of up to 17 lbs. The mass of the tips and
handle are:
Product
Code
53-1661R
53-1662R
53-1663R
53-1664R
53-1674
53-1675
53-1676
53-1674RC
53-1675RC
53-1676RC
 Effective steam sterilization can be achieved using the following parameters.
Cycle Type
Pre-vacuum
Pre-vacuum
*These sterilization parameters are for outside the US only and should not be used within the United States.
These parameters are not validated for use with FDA-cleared sterilizers.
LCN 206835-001 Rev I
®
RapidClean
1 mm Reusable Tip for the Symmetry Sharp Kerrison
2 mm Reusable Tip for the Symmetry Sharp Kerrison
3 mm Reusable Tip for the Symmetry Sharp Kerrison
4 mm Reusable Tip for the Symmetry Sharp Kerrison
Symmetry Sharp Kerrison
Symmetry Sharp Kerrison
Symmetry Sharp Kerrison
Symmetry Sharp Kerrison
®
Symmetry Sharp Kerrison
®
®
Symmetry Sharp Kerrison
Sterilization Parameters
Minimum
Temperature
132ºC/270ºF
134ºC/273ºF*
®
Handle.
®
Handle can be sterilized in the closed/assembled position.
Description
®
Rongeur Handle, for 7 inch Kerrison
®
Rongeur Handle, for 8 inch Kerrison
®
Rongeur Handle, for 9 inch Kerrison
RapidClean
®
Handle, for 7 inch Kerrison
RapidClean
®
Handle, for 8 inch Kerrison
®
RapidClean
Handle, for 9 inch Kerrison
Minimum
Exposure Time
4 minutes
3 minutes
®
Handle and the
Mass (oz)
®
Handle
0.2  0.1
®
Handle
0.4  0.1
®
0.4  0.1
Handle
0.5  0.1
®
Handle
6.5  0.2
7.0  0.2
7.4  0.2
6.5  0.2
7.1  0.2
7.6  0.2
Drying Time
30 minutes
30 minutes