CONNECTIONS
AC power inlet
AC power switch
AC power inlet
Fuse
Before first use, ensure that the voltage requirements of the PCR Workstation exactly match your
local AC power supply. Connect the PCR Workstation to the AC power supply only by the AC power
cord delivered with your PCR Workstation. Only connect the PCR Workstation to electric power
systems with an earth conductor. The AC power inlet and the AC power switch are located at the rear
of the workstation. The AC power switch serves as a disconnection of the AC power supply from the
device. The accessibility of the AC power switch must be not hindered or obscured in any way.
The electric fuse is placed at the back of the device between the power switch and the AC power
inlet. After pulling off the AC power cord and pushing upwards a safety catch, the fuse can be
changed. Use only a fuse with correct values (see "SPECIFICATIONS", page 10). Indications about
the fuse type are also located at the back of the device.
If a service is required, contact your local distributor or the manufacturer (see "TECHNICAL
SERVICE AND ORDER INFORMATION", page 27).
The serial number of the PCR Workstation is located at the rear of the instrument. The replacement
of single components (except polycarbonate panels, UV tubes and the UV Air Recirculator fuse and
filter) must not be done by the user but exclusively by authorised specialists using original
replacement parts. Therefore, the device must be sent to the manufacturer. Unauthorised work on the
device voids the warranty.
The surface of the device should be decontaminated with a lint-free cloth soaked with 70%
ethanol before sending. Please fill in and sign the decontamination certificate and send it to
the manufacturer together with the device.
For safety reasons the device must be switched off and the mains plug pulled before cleaning
is performed!
11
VWR Collection PCR Workstation HEPA