Herunterladen Inhalt Inhalt Diese Seite drucken

Siemens SITRANS TSinsert Bedienungsanleitung Seite 14

Temperature sensors
Vorschau ausblenden Andere Handbücher für SITRANS TSinsert:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 84
Mounting positions
Note
Mounting positions for small pipe diameters
With small pipe diameters, mount the sensors upstream at an angle or in an elbow, see
positions".
The following diagram shows the possible mounting positions of the sensors:
At a right angle to the flow
In a bend upstream
In a narrow cable at an angle upstream
Figure 4-1
Mounting positions
4.3
Install SITRANS TS300 Clamp-on
Note
Measurement position
Install on circular pipes only, avoid installation next to angled pipes, near slide valves, valves, pumps, etc.
1. Define the measurement position.
2. Apply heat sink compound to the metallic part of the temperature sensor.
3. For standard version: halved pipe collar allows for quick and easy mounting on the pipe using two mounting screws.
For bracket version: mount on pipe using one mounting screw.
– If the pipe is not fully occupied by medium during installation, mount the temperature sensor on the underside of the
pipe.
4. Firmly tighten the mounting screws (4 Nm torque).
5. Mount the vibration protection and hand-tighten.
– You can remove the measuring insert only after you have released the Pt100 recessed grip screw(s).
– Do not twist the housing.
– Use Pt100 recessed grip screw for installation, only.
– Do not apply force to the transmitter housing (e. g. during opening/closing of lid).
– Because the locking plugs are fitted with internal gaskets they are only suited for ambient temperatures to 100 °C
(212 °F).
14
and
in the diagram "Mounting
SITRANS TSinsert/TS100/TS200/TS300/TS500
A5E03920348-04, 09/2013

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis