→ A scale will appear in the display of the remote control and shows the value currently set for the parameter
as well as the parameter's boundary values (see fig. 35).
3) Use the ,
keys to change the parameter setting and confirm this with the key.
To leave the menu without saving changes, use the ,
the
key.
→ The message "SUCCESSFUL" appears briefly to confirm these changes were saved in the prosthesis.
4) If necessary, switch off the remote control by selecting and confirming the menu option "RC POWER OFF".
If you need to change the settings of another mode, you must first switch into that mode.
10.3.2 Using remote control app to change prosthesis setting
1) Once the component is connected and in the desired MyMode, tap the
→ A list appears with the parameters for the currently active mode.
2) Change the setting of the desired parameter either by tapping the "<", ">" symbols or moving the slider con
trol.
If you need to change the settings of another mode, you must first switch into that mode.
10.3.3 Overview of adjustment parameters in basic mode
The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait cycle. These para
meters act as basic settings for automatically adjusting the damping behaviour to the current motion situation (e.g.
ramps, slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further information on the stance
function (see Page 80). Further information on the sitting function (see Page 81).
The following parameters can be modified:
Parameter
RESISTANCE
STANCE FUNCTION
C-Leg 3C98-3/3C88-3
Adjustment soft
Remote con
ware range
trol/app adjust
ment range
120 to 180
+/- 10
keys to select the symbol
symbol in the main menu.
Meaning
Resistance against flexion motion e.g. when
walking down stairs or when sitting down
Activation/deactivation of the stance function.
This function needs to be enabled in the
adjustment software.
Use
and confirm this with
Ottobock | 85