4.625 x 6 inches (117 mm x 152 mm)
Figure 2. Tightening the setscrews in the self-sealing grommet.

Implanting the neurostimulator

1. Place the neurostimulator into the subcutaneous pocket with the Medtronic logo facing
outward, away from muscle tissue, and ensure that the extension is not bent sharply.
#
Cautions:
Ensure that the neurostimulator is placed no deeper than 4 cm (1.5 in) below
the skin and is parallel to the skin. If the neurostimulator is too deep or is not
parallel to the skin, telemetry may be unsuccessful.
Do not coil excess extension in front of the neurostimulator. Wrap excess
extension around the perimeter (Figure 3) of the neurostimulator to minimize
subcutaneous pocket depth, help minimize potential damage during
neurostimulator replacement surgery, help minimize potential kinking of the
extension, and minimize interference with telemetry. Excess extension should
not exceed two wraps around the perimeter of the neurostimulator. Extension
lengths requiring more than two wraps can interfere with telemetry.
M948681A002
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Template version: 12-13-2011
37703 2012-03