Guidelines and MANUFACTURER-Declaration - ELECTROMAGNETIC INTERFERENCE IMMUNITY
The KaWe MedCenter® 5000 is designed to be used in the types of ELECTROMAGNETIC ENVIRONMENTS listed below.
The customer or user N13) is responsible for ensuring that this device is used in such an environment.
INTERFERENCE
IEC 60601-
IMMUNITY-TEST
TEST LEVEL
ELECTROSTATIC
DISCHARGE (ESD)
± 6kV contact discharge
according to IEC
± 8 kV air discharge
61000-4-2
Fast transient electrical
± 2kV for power supply
disturbances/ bursts
lines
according to IEC
± 1kV for input and
61000-4-4
output lines
± 1kV phase-to-
Surges according to IEC
phase voltage
61000-4-5
± 2kV phase-to-ground
voltage
<5% Uτ
for 1/2 period
(>95 % dip)
Voltage dips, short
40 % Uτ
interruptions
for 5 periods
and voltage variations
(60 % dip)
on power supply input
lines according to IEC
70% Uτ
61000-4-11
for 25 periods
(30 % dip)
<5 % Uτ for 5 ѕ
(>95 % dip)
Magnetic field with
a power frequency of
3 A/m
(50/60 Hz) according
to IEC 61000-4-8
NOTE: Uτ is the mains voltage prior to application of the test level.
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT –
LEVEL
GUIDELINES
The flooring should be wood or concrete or be covered
± 6kV contact
with ceramic tile. If the floor material is a non-conduc-
discharge
tive, synthetic material, the relative humidity of the air
± 8 kV air discharge
must be at least 30%
± 2kV for power supply
The quality of the mains supply voltage should be
lines
the same as that of a typical commercial or hospital
± 1kV for input and
environment.
output lines
± 1kV phase-to-
The quality of the mains supply voltage should be
phase voltage
the same as that of a typical commercial or hospital
± 2kV phase-to-
environment.
ground voltage
<5% Uτ
for 1/2 period
(>95 % dip)
The quality of the mains supply voltage should be
40 % Uτ
the same as that of a typical commercial or hospital
for 5 periods
environment.
(60 % dip)
If the user requires continued OPERATION during power
interruptions, it is recommended that the device be
70% Uτ
powered from an uninterruptable power supply or a
for 25 periods
battery.
(30 % dip)
<5 % Uτ for 5 ѕ
(>95 % dip)
Should interference occur, it may be necessary to place
the KaWe MedCenter® 5000 further from the source
line frequency magnetic fields or to fit it with a magne-
3 A/m
tic shield: the line frequency magnetic field should be
measured at the intended use site to ensure that it is
not large enough to cause a disturbance.