Herunterladen Inhalt Inhalt Diese Seite drucken

Widex CALL-DEX Bedienungsanleitung Seite 306

Vorschau ausblenden Andere Handbücher für CALL-DEX:
Inhaltsverzeichnis

Werbung

Electromagnetic immunity
The CALL-DEX is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of a CALL-DEX should assure that it is
used in such an environment.
Immunity
Test
Electro-
static
discharge
(ESD)
IEC
61000-4-4
Electrical
fast
transients/
burst
IEC
61000-4-5
Surge
IEC
61000-4-5
IEC 60601
Test level
± 6 kV
contact
± 8 kV air
± 2 kV for
power line
supplies
± 1 kV for
input/
output lines
± 1 kV line(s)
to line(s)
± 2 kV
line(s) to
earth
Compliance
level
± 6 kV
contact
± 8 kV air
Not
applicable *)
Not
applicable *)
Not
applicable *)
Not
applicable *)
304
Electromagnetic
environment
– guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Not
applicable *)
Not
applicable *)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis