TB2 and TB4 – Terminal Assignment for Oxygen, Conductivity, pH and Ozone – ISM Sensors
TB2
(ISM Ch1,2)
Term. Function
1
DI2+
2
DI2–
3
1-Wire_Ch1
4
GND5V_Ch1
5
RS485B_Ch1
6
RS485A_Ch1
7
GND5V_Ch1
8
5V_Ch1
9
24V_Ch2
10
GND24V_Ch2
11
1-Wire_Ch2
12
GND5V_Ch2
13
RS485B_Ch2
14
RS485A_Ch2
15
GND5V_Ch2
16
5V_Ch2
TB3 – Terminal Assignment for Flow Sensors
TB3
Term. Transmitter
9
Ain_Ch5
10
AJ_Ch5
11
5V_Ch5
12
GND5V_Ch5
13
Ain_Ch6
14
AJ_Ch6
15
5V_Ch6
16
GND5V_Ch6
TB4
Optical Oxygen
(ISM Ch3,4)
Function
VP8 cables:
wire color
DI6+
–
DI6–
–
1-Wire_Ch3
–
GND5V_Ch3
–
RS485B_Ch3
–
RS485A_Ch3
–
GND5V_Ch3
–
5V_Ch3
–
24V_Ch4
grey
GND24V_Ch4
blue
1-Wire_Ch4
–
GND5V_Ch4
green/yellow grey
RS485B_Ch4
brown
RS485A_Ch4
pink
GND5V_Ch4
–
5V_Ch4
–
Flow hi, Flow lo Flow Type2
Function
Flow Pulse Input
+10 VDC
+5 VDC
Ground
Flow Pulse Input
+10 VDC
+5 VDC
Ground
, CO2 hi
UniCond 2e
1)
1)
5-pin cables:
Cables:
wire color
wire color
–
–
–
–
–
–
–
–
–
black
–
red
–
white
–
blue
brown
–
black
–
–
–
–
blue
black
white
red
yellow
white
–
blue
1) Always one O2 optical or CO2 hi
sensor can be connected to plug TB2
and TB4.
2) Transparent wire not connected.
pH, Amp. Oxygen, CO2,
2)
Cond 4e and O3
Cables:
wire color
–
–
transparent
(cable core)
red
–
–
–
–
–
transparent
(cable core)
red
–
–
–
–
11