Table of contents
Specifications
on page 3
Startup
on page 20
Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Enclosure
Weight
Pollution degree
Over voltage category
Protection class
Power requirements
Fuse
Operating temperature
Storage temperature
Humidity
Altitude
connections
Network connections
Relay card (optional)
Analog output card (optional)
Certification
Warranty
1
The AC power outlets only supply power when the instrument has the optional 100 to 240 VAC
power supply.
Details
Controller: 315 x 120 x 242 mm (12.28 x 4.68 x 9.5 in.)
USB box: 79.5 x 55.1 x 159.5 mm (3.13 x 2.17 x 6.28 in.)
Controller: Metal with corrosion-resistant surface, NEMA 4X/IP65 rating
USB box: ABS/polycarbonate, NEMA 4X/IP65 rating
Approximately 5 kg (11 lb). Weight varies by model.
2
II
I
100 to 240 VAC ± 10 VAC, 50/60 Hz, 1000 VA maximum
F1 and F2: M 3.5 A L, 250 V or T 3.15 A L, 250 V; F3 and F4: T 8 A H, 250 V
-20 to 55 °C (-4 to 131 °F)
-20 to 70 °C (-4 to 158 °F)
95% relative humidity, non-condensing
2000 m (6561 ft)
Two, four or six device connectors and two AC power outlets
Two Ethernet connectors (10/100 Mbps), switch function, M12 female D-coding
connector
One USB connector in a USB box
Four relays on each relay card, change over contacts (SPDT)
Maximum switching voltage: 250 VAC, 125 VDC
Maximum switching current: 5 A
Note: Make sure to install the 5 A external breaker.
Maximum switching power: 1500 VA, 250 VAC; 625 W, 125 VDC
2
Wire gauge: 1.5 mm
(15 AWG) maximum
Four 4-20 mA analog outputs on each analog output card, 500 Ω maximum
2
Wire gauge: 1.5 mm
(15 AWG) maximum
The manufacturer recommends that signal cables with a shield are used.
cTUVus compliant, CE compliant, DIN EN 61326 surge protection
1 year (EU: 2 years)
Operation
on page 20
Maintenance
on page 20
1
English 3