Herunterladen Inhalt Inhalt Diese Seite drucken

Temperature Measurements; Mv Measurements; Saving, Viewing And Deleting Data - VWR pH110 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
3. If AUTOLOCK is off, the pH value of the sample displays on the screen. If both pH and
temperature readings are stable, take a reading. If AUTOLOCK is on, press MEA./EFF.
"WAIT" flashes until the unit determines a stable pH reading.

Temperature Measurements

The VWR pH110 can measure temperature independently with the ATC/Temp probe without
using the pH electrode. Place the ATC/Temp probe in the media to be measured. The
measured temperature displays.

mV Measurements

1.
Connect the optional combination mV electrode to the unit. Press MODE to enter mV mode
with AUTOLOCK on or off as desired. For inherently unstable samples, the unit will not
AUTOLOCK. Use mV mode with AUTOLOCK off in this case.
2.
Rinse electrode with distilled water and immerse it in sample to be measured. If AUTOLOCK
is off, the mV value of the sample will be displayed on the screen. If AUTOLOCK is on, press
MEA./EFF. "WAIT" flashes until the unit determines a stable mV reading.

Saving, Viewing and Deleting Data

The pH110 can save 50 data records. When in measurement mode, press
record. The instrument will confirm the saved data by displaying SAVE and the record number
for one second. "Full" is displayed when trying to save data and the memory is full.
To view saved data, press mode until RECALL is displayed and then press
Up or Down arrow keys to review different saved records. Press Mode to escape back to
measurement mode.
To delete data records, press Mode while in measurement mode until DELETE is displayed.
Press
ESC . "All" will be displayed and blinking. Press the Up or Down arrow key to switch
between delete 'All' or 'Each' options. Select either 'All' or 'Each' by pressing
option is displayed.
If 'All' is selected, all records will be deleted from memory and 'None' will be displayed. Press
Mode twice to return to the measurement mode.
If 'Each' is selected, the Up and Down arrow keys will allow you to scroll through the saved data
records. Press
ESC to delete the selected record. All records after the deleted record will
shift up to keep the records in sequential order. For example, if record 3 is deleted, record 4 will
become record 3 and record 5 will become record 4. Press Mode twice to return to the
measurement mode.
4
ESC to save a
.
ESC
Use the
ESC while that

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis