MP6025871-12C1 / A
Disconnecting the transmitter from the sensor
Before proceeding, have your pump user guide available.
To disconnect the transmitter from the sensor:
1
Carefully remove any occlusive dressing from the
transmitter and sensor.
2
For the Enlite sensor, remove the adhesive tab from the top
of the transmitter.
3
Hold the transmitter as shown, and pinch the flexible side
arms of the sensor between your thumb and forefinger.
4
Gently pull the transmitter away from the sensor.
5
Follow the instructions that appear on the pump or follow
the instructions in your pump user guide.
Removing the sensor
Always refer to the sensor user guide for instructions on how to remove the sensor.
Bathing and swimming
After the transmitter and sensor are connected, they form a waterproof seal to a
depth of 2.4 meters (eight feet) for up to 30 minutes. You can shower and swim
without removing them. No occlusive dressing or overtape is needed.
Watertight Tester
The tester is used to test the transmitter to make sure it is working. It is also used as
a required component for cleaning the transmitter. Properly connecting the tester to
the transmitter will ensure that fluids do not come in contact with the transmitter's
connector pins. Fluids can cause connector pins to corrode and affect the
transmitter's performance.
Do not twist the tester while attached to the transmitter. This will damage the
transmitter.
The tester can be used for one year. If you continue to use the tester for more than
one year, the transmitter's connector pins could be damaged, because the tester
cannot continue to provide a waterproof seal. For instructions on how to check the
connector pins, see Inspecting the transmitter connector pins, on page 8.
-7-