5. Commissioning, operation
5. Commissioning, operation
Personnel: Skilled personnel
Tools: Spanner or torque spanner
Only use original parts (see chapter 10 "Accessories").
DANGER!
Danger to life from explosion!
Through working in flammable atmospheres, there is a risk of explosion
which can cause death.
Only carry out set-up work in non-hazardous environments!
WARNING!
Physical injuries and damage to property and the environment caused
by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.

5.1 Mounting the instrument

Only use the pressure transmitter if it is in perfect condition with respect to safety.
Prior to commissioning, the pressure transmitter must be subjected to a visual inspection.
Leaking fluid is indicative of damage.

5.2 Requirements for mounting point

The mounting point must meet the following conditions:
Ambience corresponds to a max. pollution degree of 3.
Humidity ≤ 80 % r. h. (non-condensing).
Protected from weather influences.
Sealing faces are clean and undamaged.
Sufficient space for a safe electrical installation.
For information on tapped holes and welding sockets, see Technical information
IN 00.14 at www.wika.com.
Permissible ambient and medium temperatures remain within the performance limits.
Consider possible restrictions on the ambient temperature range caused by mating
connector used.
For performance limits, see chapter 9 "Specifications"
WIKA operating instructions, reference pressure sensor, model CPT62I0
EN
11