Initial operation
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety margin
of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specifications".
Turn on the mains switch (Switch position "").
The machine type and program version will be displayed and the LEDs light up.
displayed after 15 seconds.
Open the lid.
The last used centrifuge data will be displayed.
Remove the transport safety device (see instruction sheet on "Moving the equipment safely").
9
Opening and closing the lid
9.1
Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Rotate the turning handle on the front panel to the left.
Open the lid.
9.2
Closing the lid
Do not bang the lid shut.
Place the lid and lightly press down the front edge of the lid.
Turn the handle on the front panel to the right.
10
Installation and removal of the rotor
A
B
C
D
36/93
Rotor installation:
Dirt particles between the motor shaft and rotor prevent the rotor from
having a perfect seat and cause it to run unsteadily.
Clean the motor shaft (C) and the bore of the rotor (A) and then apply a thin coat
of grease to the motor shaft.
Place the rotor vertically onto the motor shaft. The driver of the motor shaft (D)
must be in the groove of the rotor (B). The groove alignment is marked on the
rotor.
Tighten the clamping nut of the rotor by turning it in the clockwise direction
using the key included in delivery.
Check the rotor to make sure it is seated firmly.
Removing the rotor:
Loosen the clamping nut by turning it in the counter clockwise direction up to the
lifting pressure point. After overcoming the lifting pressure point, the rotor is
released from the cone of the motor shaft.
Turn the clamping nut until the rotor can be lifted up from the motor shaft.
OPEN
OEFFNEN will be