Mechanical installation
Version
Standard
Rustless
1)
Recommended value for the specified property class. Please comply with the screw
manufacturer's instructions with regard to screw dimensions.
Dimensions in mm (1 mm = 0.03937 inches)
Installation: 2 x M6
tightening torque 10 Nm
max. thread reach
Fig. 4.1
Load application and installation
Important
Load must not be applied to the side where the cable connection is located, as
this would cause a force shunt.
10
Thread
Max. thread
reach
M6
10 mm
M6
10 mm
Intermediate plate
(customer construction)
10 mm
Platform
(customer construction)
A03627_02_Y00_01 HBM: public
Min. property
A2-70 or A4-70
Load direction
Mounting base
Intermediate plate
Tightening
class
8.8
Installation: 2 x M6
tightening torque 10 Nm
max. thread reach 10 mm
1)
torque
10 N⋅m
10 N⋅m
Connector
plug
PW25P