Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Specifications - Beurer BM 54 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für BM 54:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

11. Technical specifications

Model No.
BM 54
Measurement me-
Oscillometric, non-invasive blood pres-
thod
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 250 mmHg,
Systolic 50 – 250 mmHg,
Diastolic 30 – 200 mmHg,
Pulse 40 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 139 mm x W 94 mm x H 48 mm
Weight
349 g (without batteries, with cuff)
Cuff size
22 to 44 cm
Permissible operating
+ 5°C to + 40°C, 15-93% relative humi-
conditions
dity (non-condensing)
Permissible storage
-25°C to + 70°C, ≤ 93% relative humidity,
700 – 1060 hPa ambient pressure
conditions
Power supply
4x 1.5 V
AAA batteries
Battery life
Accessories
Classification
Data transfer via
Bluetooth
wireless
®
technology
The serial number is located on the device or in the battery com-
partment.
Technical information is subject to change without notification
to allow for updates.
• This device complies with European Standard EN60601-1-2
and is subject to particular precautions with regard to electro-
magnetic compatibility. Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Services
address or found at the end of the instructions for use.
• The device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinproduk-
tgesetz) and the standards EN1060-1 (non-invasive sphyg-
momanometers – Part 1: General requirements), EN1060-3
30
Lasts for approx. 200 measurements
depending on the blood pressure and
pump pressure as well as the number
of Bluetooth
connections.
®
Cuff, instructions for use, storage pouch
Internal supply, IP21, no AP or APG,
continuous operation, application part
type BF
The blood pressure monitor uses
Bluetooth
low energy technology,
®
2402MHz – 2480MHz frequency band,
Transmission power max. 20 dBM,
Compatible with Bluetooth 4.0 smart-
phones/tablets

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis