11 Annual Product
Inspection
(Therapist, Technician, Leckey Product
Advisor, Dealer)
Leckey recommend that each product should
be subject to a detailed inspection at least
once a year and every time the product is
reissued for use. This inspection should
be carried out by a technically competent
person who has been trained in the use of
the product and should include the following
checks as a minimum requirement.
1. Check all knobs, nuts and bolts are in
place, replacing any missing items.
2. Check the chassis height adjustment
mechanism is working properly. If the
chassis is foot pedal operated ensure the
seat height doesn't change when the pedal
is released. Also, check that the locking
pin engages securely to prevent accidental
height adjustment of the chassis.
3. Adjust the seat to its maximum range of
tilt in space and ensure that the locking lever
locks the seat out securely at varying point in
this range.
4. Check that where the seat and chassis
join there is no visible wear and tear on the
metal components.
5. Lift the base to check each castor
individually. Make sure they are moving
freely and remove any dirt from the rubber
wheels. Check that the brakes lock the
wheels securely.
6. Visually check the structure of the
product paying attention to weld points
on the frame ensuring there are no signs
of fatigue or cracking around the welds.
7. Leckey recommend that a written
record is maintained of all annual
product inspections.
If in any doubt to the continued safe
use of your Leckey product or if any
parts should fail, please cease using
the product and contact our customer
service department or your local dealer
as soon as possible.
12 Re-issuing Leckey
Products
Most Leckey products are assessed and
ordered to meet the needs of an individual
user. Before reissuing a product we
recommend that the therapist prescribing
the product has carried out an equipment
compatibility check for the new user and
has ensured that the product being
re-issued contains no modifications or
special attachments.
A detailed technical inspection should be
carried out on the product prior to re-issuing.
This should be carried out by a technically
competent person who has been trained
in the use and inspection of the product.
Please refer to section 11 for the required
checks to be carried out.
Ensure the product has been cleaned
thoroughly in accordance with section 9 of
this manual.
Ensure a copy of the user manual is supplied
with the product. A copy can be downloaded
from our website www.leckey.com
Leckey recommend that a written record is
maintained of all product inspections carried out
during the reissue of the product.
If in any doubt to the continued safe
use of your Leckey product or if any
parts should fail, please cease using
the product and contact our customer
service department or your local dealer
as soon as possible.
13 Product Servicing
Servicing of all Leckey products should only
be carried out by technically competent
persons who have been trained in the use of
the product.
In the UK & ROI please contact the Leckey
Service Centre on UK 0800 318265 or ROI 1800
626020 and our customer service department
will be delighted to assist you with your
servicing requirements.