CELSIUS H270 MED and Components
Information for ensuring electromagnetic compatibility
●
All data and signal cables must have sufficient shielding. Requirements EN 50173 and
EN 50174-1/2 apply to LAN cabling. Use of unshielded or badly shielded cables may lead to
increased emission of interference and/or reduced fault-tolerance of the device.
●
All casing covers must be properly secured.
●
Portable and mobile wireless communications devices such as mobile phones may influence
medical electrical devices. It is imperative that you observe the safety distances given in
table 4.
●
Protect the contacts of all sockets and plugs of the device against static electricity. Avoid
touching the contacts. Should touching be unavoidable, take the following safety measures:
–
Touch an earthed object before touching the contacts. This discharges static charges.
or
–
Wear an earthing strap.
●
Use only the cables supplied when connecting peripheral devices, such as monitors.
●
Each peripheral device you want to connect must comply with the requirements for use in an
environment with a medical application.
●
Install only system expansions that satisfy the requirements and rules governing safety and
electromagnetic compatibility and relating to telecommunications terminal equipment. If you
install other expansions, you may damage the device or violate the safety regulations and
regulations governing RFI suppression.
The effects of static electricity, fast transients or strong electromagnetic fields on the
i
device can cause malfunctions. During the malfunction the picture and sound quality, as
well as data communication with USB devices may be impaired.
When operating the analogue modem connection, reduced data transfer rates, as well as
aborting of the data transfer or the data connection can result due to the effects of induced
high-frequency fields.
2 - English
A26391-K270-Z290-1-8N19, edition 1