Input of of temperature (Manual temperature compensation)
The "PARA TEMP" submenu facilitates the input of temperature values for pH measurements
without temperature probe. The default value is 15°C.
This submenu only appears if a manual temperature compensation (CONF T-C: T C MANUAL)
was selected in the configuration level.
Setting:
126
Parameterization
Submenu "PARA TEMP".
With
enter submenu
"PARA TEMP".
Default value 15°C.
With
set temperaure value between
0°C and 50°C.
With
accept and return to submenu
"PARAMETER".