6
List of tools for Prosonic Flow 92 F, Prowirl 72, 73
3 mm, 4 mm,
5 mm, 6 mm
7
List of tools for Prowirl 200
3 mm, 4 mm,
5 mm, 6 mm
8
Important information regarding the replacement of
the compact and remote sensor version
1. The sealing surface may not be scratched.
2. Remove the sealing washer with a suitable tool without scratching the sealing surface.
3. Clean the sealing surface on the meter body with a suitable solvent and a lint-free cloth.
4. The bore must be perfectly clean.
5. Place the new sensor seal on the sealing surface with the printed surface (if present) facing
upwards (see example in the graphic below).
1
‣
Example: sensor seal with printed
surface
6. Grease the threads and the head contact areas of the sensor screws.
Endress+Hauser
8 mm, 17 mm
8 mm, 17 mm
0.5 × 3.5 mm
T10, T20
0.5 × 3.5 mm
T10, T20
EA01085D
Acid-free, non-
hardening lubricant
9