CAUTION: The operator should not touch the contacts of the battery
and the patient at the same time.
Please Note: Dispose of the waste battery in accordance with EU
Waste Battery Regulations.

Cleaning Instructions

Disinfection of contaminated parts is only effective after careful
preliminary cleaning. CareFusion recommends the tested solution of
PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be wiped with any aqueous solutions and must
not be exposed to solvents e.g. alcohol, chloride solutions, as there
are electronic components inside that will be permanently damaged.
CAUTION: Switch off the device before cleaning.

External Surfaces of the Spirometer

CAUTION: Do not attempt to wash or immerse the Respiratory
Pressure Meter in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
The external housing of the pressure meter may be wiped with sterile
wipes or a damp cloth that has been immersed in a cold sterilizing
solution.
41