Herunterladen Inhalt Inhalt Diese Seite drucken

Caire Eclipse with autoSat Bedienungsanleitung Seite 13

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 44
Typical New Power Cartridge Operating Times
Flow Rate
Continuous
Setting
Flow (LPM)
0.5
4.4 hours
1.0
3.7 hours
2.0
2.0 hours
3.0
1.3 hours
Note: Operating the Oxygen Concentrator outside of its normal operating
temperature range can affect performance and decrease battery run time
and/or increase battery charge time. (Refer to the Specifications section in
this manual.) You will achieve longer operating time on the Battery if you
operate your Eclipse in the Pulse Mode (Physician approved mode).
mL
Pulse Dose Mode
(12 BPM)
SETTING
-
-
16mL 1.0
5.4 hours
32mL 2.0
5.1 hours
48mL 3.0
4.9 hours
64mL 4.0
4.0 hours
80mL 5.0
3.7 hours
96mL 6.0
3.5 hours
128mL
2.5 hours
160mL
2.0 hours
192mL
1.7 hours
Typical Battery Recharge Time
The typical time to recharge your Battery to achieve 80%
capacity from a fully discharged Battery is between 2 and 5
hours, depending upon the device flow setting. When the battery
is fully charged, the battery gauge will cease cascading. The
battery can be charged in the unit when the Eclipse is turned off
and plugged in.
If the Battery becomes too warm during discharging, recharging
will not begin until the Battery sufficiently cools. Removing the
Battery and allowing it to cool may expedite this cooling process.
The Battery will not recharge when operating from the DC Power
Supply.
The Battery charges when in the device is operating on AC power
only. The Battery will also charge from the optional Desktop
Charger.
Eclipse User Manual — 13

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis