TROUBLESHOOTING

Type of problem
Water comes into the powder chamber
or leaks from the cap
No powder/air jet coming from the unit ‡
Air and/or powder is leaking from the
threads of the cap
7KH XQLW HI¿FLHQF\ GLPLQLVKHV
Water is leaking between the handpiece
and the unit
Air is leaking between the handpiece
and the unit
Air and/or water is leaking from the
connector of the cord
74
Solutions
Check the dental unit connection
Check the state of the o-rings of the turbine connection
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
'U\ DQG ¿OO WKH SRZGHU FKDPEHU
Send the complete unit to an approved EMS repair center
Cut immediately the air input by releasing the foot pedal
Wait for 1-2 minutes for the system to decompress
Disconnect the unit from the dental unit
Press the foot control pedal of the dental unit
If no air comes out of the turbine connector, it is your dental unit that causes the problem
If air comes out of the turbine connector, it is the device that causes the problem
If air comes through the handpiece connector, the handpiece is blocked.
Clean the handpiece with the EasyClean
Unscrew the cap above the wash-hand basin, because the powder can be ejected.
A sealed apparatus, even disconnected, can indeed remain under pressure
Empty the powder chamber and screw the cap back on
Connect the unit to the dental unit (be careful, the powder chamber must be empty)
Press the foot control pedal of the dental unit
If no air comes out of the handpiece connector, the unit is blocked
Send the complete unit to an approved EMS repair center
Check the seal and the cleanliness of the thread on the powder chamber and on the cap
Replace the seal or the cap if necessary
6HQG WKH GHYLFH WR D FHUWL¿HG (06 5HSDLU &HQWHU
$ SRZGHU UH¿OO LV SHUKDSV QHFHVVDU\
Clean the handpiece
6HQG WKH GHYLFH WR D FHUWL¿HG (06 5HSDLU &HQWHU
Replace the seal of the handpiece connector
Send the complete unit to an approved EMS repair center
Clean the handpiece
Replace the seal of the handpiece connector
Send the complete unit to an approved EMS repair center
Check the connector of the cord, if it is necessary, carry out maintenance following the information of the
manufacturer
Send the complete unit to an approved EMS repair center