6. Use CLEAR, ⇑, ⇓ and shift to input full tank weight.
7. Press ENTER to save the full tank weight. The display will read "GOOD" to confirm.
8. Press ON/OFF button on the monitor and platform to turn unit off.

CALIBRATION PROCEDURE

1. Turn on platform.
2. Turn on monitor. Press ⇑ and ⇓ together. Display will then show "SEt-1."
3. Press TARE/ENTER. Display will show "SEt-11" for kg calibration (for lb calibration press ⇑ and
"SEt-12" will display.) Press TARE/ENTER to confirm.
4. A "-----" 5 digit number will be displayed, press ⇓.
5. Place calibration weight on scale.
6. Press ⇑. Press TARE/ENTER. Number displayed must match calibration weight. If a digit needs
to be changed, press UNITS/SHIFT to activate digit (digit will blink.) Press the ⇑ or ⇓ to change
the number. Repeat until all digits are changed.
7. When the number displayed agrees with the weight on the scale press TARE/ENTER.
Calibration is complete. Monitor will restart automatically.

SPECIFICATIONS:

• Capacity: 243 lbs (110 kg)
• Accuracy: [±0.01 lb / ±0.005 kg] or 0.1% of reading whichever is greater.
• Resolution: 0.02 lb / 0.01 kg
• Automatic Shut-Off: 20 minutes
• Operating Temperature: 32˚F to 120˚F (0˚C to 49˚C)
• Battery Life: Monitor - approximately 20 hours; platform - approximately 30 hours

REPLACING THE BATTERY:

Replace the 9V battery in monitor or (6) AA batteries in platform when low-battery symbols
prompt on LCD. DO NOT OPERATE WITH A LOW BATTERY.