Warning
Description
PH MAINT RECD
pH sensor maintenance is recommended.
The offset is outside of ±45 mV but within ±60 mV.
T MAINT RECD
The temperature offset is outside of ±3 °C but within
±5 °C.
Replacement parts and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Parts
Description
Cable, gateway to controller, 1 m
Cable assembly, 4 pin, shielded, flow sensor, 2 m
Cable assembly, 5 pin, shielded, M12, chlorine sensor, 1 m
Fitting, inlet port for chlorine flow cell
Fitting, speed fit, outlet port for chlorine flow cell or inlet port for pH flow cell
Fitting, elbow, outlet port for pH flow cell
Float, flow sensor
Flow cell assembly, chlorine
Flow cell assembly, pH combination
Flow cell assembly, pHD
includes sealing hub
Flow cell body only (pH)
Flow cell block, chlorine
Flow control valve
Flow control valve kit
Flow sensor
Table 3 Warning list for the sensor (continued)
W A R N I N G
Resolution
Clean the sensor and replace the
salt bridge and standard cell
solution and then repeat the
calibration, or replace the sensor.
Calibrate the temperature.
Item no.
6122400
9160900
9160300
English:
9159200 (1/4 in.)
Metric:
9196400 (6 mm)
English:
9159300 (3/8 in.)
Metric:
9196500 (10 mm)
English:
9196900 (1/2 in.)
Metric:
9197000 (12 mm)
9159100
English: 9198400
Metric: 9198500
English: 9180200
Metric: 9180400
English: 9180100
Metric: 9180300
9159700
9158900
9159000
9198600
9160800
English 19