Herunterladen Inhalt Inhalt Diese Seite drucken

Chauvin Arnoux C.A 6545 Bedienungsanleitung Seite 73

Megohmmeter
Vorschau ausblenden Andere Handbücher für C.A 6545:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 82
 Typical sttling time for the measurement according to components tested (U
These values include the infl uences of the charging of the capacitive component, of the automatic
range system and to the test voltage control.
Test voltage
500 V
1000 V
2500 V
5000 V
 Typical discharge time for a capacitive component to reach 25 V
Initial voltage
Discharge time (C in µF)
 Typical change curve for test voltages according to load
V
600
500
400
300
200
100
0
0,01
Non capacitive
Load
(unsmoothed measurement)
1 M 
100 G 
1 M 
100 G 
3 M 
100 G 
5 M 
100 G 
500 V
C x 3 s
Range 500 V
0,1
73
With capacitance of 1 µF
(smoothed measurement)
3 s
8 s
3 s
8 s
3 s
8 s
4 s
8 s
DC
1000 V
2500 V
C x 4 s
C x 4 s
= 0.03 U
)
dist
n
4s
40 s
4 s
80 s
4 s
90 s
16 s
120 s
5000 V
C x 7 s
M 
1

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

C.a 6547

Inhaltsverzeichnis