Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Ultrasound
Frequencies
Display
Contact control threshold
Timer
Ultrasound, continuous
Pulse frequency / duty cycle
Number of connections
Intensity
Ultrasound, pulsed
Pulse frequency / duty cycle
Number of connections : 1
Intensity
Treatment heads
1 and 3 MHz, large (standard)
1 and 3 MHz, small (option)
*SATP = Spatial Average Temporal Peak (average pulse power)
*ERA = Effective Radiating Area, this is the effective radiating area of the treatment head.
*BNR = Beam Nonuniformity Ratio, indicates the ratio between the peaks and the average value of the
intensity in the sound beam. A low BNR excludes high, unwanted energy concentrations.
ERA and BNR have been measured according the FDA-method. All treatments heads have been provided
with contact control.
Current general
Number of channels
Output characteristics
Resolution of CV or CC display
Timer
The maximum amplitude is achieved within the specification at a load of max. 500 Ohm (CC).
Two-pole interferential therapy
Carrier frequency
Stimulation frequency (AMF)
Frequency modulation (spectrum)
Frequency modulation programs
Amplitude
Asymmetrical biphasic pulsed current
Phase duration
Frequency
Burst frequency
Frequency modulation (spectrum)
Frequency modulation programs
Amplitude
Medium-frequency direct current
Frequency
Duty cycle
Amplitude
Programming
Memory
: 1 and 3.1 MHz
: intensity in W/cm² and power output in W (SATP*)
: 80%
: 0 - 30 minutes, coupled to contact control
: 100 Hz / 100 %
: 1
: 0 - 1.5 W/cm²
: 100 Hz / 5, 10, 20, 50, 80 %
: 0 - 3 W/cm², duty cycle 5, 10, 20, 50 %
0 - 1.8 W/cm², duty cycle 80 %
: Geometric surface 5.8 cm
side panel radiation max. 10 W/cm² (IEC requirement < 100 mW/cm²)
: Geometric surface 1.15 cm
side panel radiation max.10 mW/cm² (IEC requirement < 100 mW/cm²)
: 1
: Constant Current (CC) or Constant Voltage (CV)
: in steps of 1 mA
: 0 - 30 min.
: 2, 4 or 10 kHz
: 0-5-10-15-20-30-40-50-80-100-150 Hz
: 0-10-20-30-40-50-80-100 Hz
: 6/6s and 1/30/1/30s
: 0 - 100 mA
0 - 60 mA in combination therapy
: 20-30-40-50-75-100-125-150-175-200-300-400 µs
: 5-10-15-20-30-40-50-60-80-100-120-140-160-180-200 Hz
: 1, 2, 3, 4, 5, 6, 7, 8, 9 Hz
: 0-5-10-20-30-40-50-60-80-100-120-140-160-180 Hz
: 6/6s and 1/30/1/30s
: 0 -140 mA, if phase duration < 200 µs
0 - 100 mA, if phase duration is 300 µs
0 - 70 mA, if phase duration is 400 µs
: 8000 Hz
: 95%
: 0-40 mA
: memory for 10 sets of freely selectable therapy parameters

SPECIFICATIONS

2
ERA* 5 cm
2,
BNR* max. 5 type collimating,
2
, ERA* 0.8 cm
2
, BNR* max. 5 type collimating,
29

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Sonopuls 490

Inhaltsverzeichnis