CAL 1
characteristic
is shifted
11 % r.H.
CAL 2
characteristic
is turned
B - 15
offset input
measured temperature value
offset input
measured humidity value
Operating manual – English
CAL 2: The offset for sensor 2 (relative humi-
dity) is set with CAL 2 (one-point calibration).
The offset turns the characteristic around the
lower calibration point (11 % r.H.). The calibra-
tion point must be in the range from 30 % to
%r.H.
dp°C dp°F
95 % r.H. The offset is displayed in the upper
display. A maximum ± 10 % r.H. can be set.
%r.H.
dp°C dp°F
CAL 2 can only be selected in connection
CAL 2
with the unit % r.H. and at a prevailing
TH
IR
DP
CFG
humidity of at least 30 % r.H.
Selection can be made with
The factory settings are obtained by setting the offset to 0.0.
and
, confirmation with
.