Setting the emissivity
The emission level can be set from 0.10 to 1.00 individually or via an integrated material table with common metal
surfaces. Thus, different materials and surfaces can be measured with exact results.
Turn the measuring device on by briefly pressing the measuring button (5).
Within the preset power-off time, enter the "Unit/Mem/E" setting mode with the "MODE" button.
Press "F3" to access the settings menu "E".
Setting the emissivity individually
The emission level can be set between 0.10 to 1.00 in intervals of 0.01 with "F1" and "F3".
Press the measuring button (5) or "MODE" to confirm your input.
Setting the emissivity per table
The emission level can be set using an integrated material table for common metals.
Press "F2" to select the "Tab" function. A material table is displayed.
Select the material at hand with "F1" and "F3". The selected material is highlighted with a bar. The emission level
changes accordingly.
Default
Emission value of most materials (0.95)
Ox Aluminum
Aluminium oxide (0.30)
Ox Brass
Brass oxide (0.50)
Ox Copper
Copper oxide (0.60)
Paint
Painted surface (0.93)
43