Technical data

Mode of operation
Measuring range > 100 to
< 600 mbar
Input
Measured variable input
Measuring range for
Overload limit
gauge pressure
0 ... 100 mbar g
> -400 / < 400 mbar g
0 ... 160 mbar g
> -400 / < 400 mbar g
0 ... 250 mbar g
> -800 / < 1 000 mbar g
0 ... 400 mbar g
> -800 / < 1 000 mbar g
0 ... 600 mbar g
> -1 000 / < 2 000 mbar g
Output
Current signal
Burden
Auxiliary power U
B
Current consumption I
B
Voltage signal
Burden
Auxiliary power U
B
Current consumption
Characteristic
Piezoresistive with stainless steel
diaphragm
Burst pressure
1 bar g
1 bar g
2 bar g
2 bar g
3 bar g
4 ... 20 mA
(U
- 10 V) / 0.02 A
B
7 ... 33 V DC
(10 to 30 V for
hazardous areas)
< 20 mA
0 ... 10 VDC
> 10 kΩ
12 ... 33 VDC
< 7 mA at 10 kΩ
Linear rising
Measuring accuracy
Measurement deviation at 25 ÀC (77 ÀF),
Characteristic deviation, hysteresis and
repeatability included
Setting T99
Long-term drift
Start-of-scale value
and measuring span
Ambient temperature influence
Start-of-scale value and
measuring span
Vibration influence ( complying
with IEC 60068-2-6)
Auxiliary power influence
Conditions during operation
Ambient conditions
Ambient air temperature
Altitude
Relative humidity
Storage temperature
Degree of protection
(complying with EN 60529)
Electromagnetic compatibilityª
Construction
Weight
Process connections
Electrical connections
Material of the parts in contact with measured material
Measuring cell
stainless steel , material no. 1.4435
Process connection
stainless steel, material no. 1.4404 (SST 316 L)
Sealing material
Position 15 of
order number
Viton (FPM)
A
Neoprene (CR)
B
Perbunan (NBR)
C
EPDM
D
Material of parts not in contact with the medium
Housing
stainless steel, material no. 1.4404 (SST 316 L)
Pin and socket
plastic
connector housing
CuZn, nickel-plated (plug M12)
Cable
PVC spec.
typically: 0.25 % of full scale
value
maximum: 0.5 % of full scale
value
< 0.1 s
0.25 % of full scale value/year
0.25 %/10 K of full-scale value
0.5 %/10 K of full-scale value
for a
measuring range 100 ... 400
mbar
0.005 %/g to 500 Hz in all
directions
0.005 %/V
Outdoor and indoor use
-25 ⁄ +85 ÀC (-13 ... +185 ÀF)
max. 2 000 m ASL
Use an appropriate power supply
for altitudes higher than 2000
m ASL.
0 ... 100 %
-50 ⁄ +100 ÀC (-58 ... +212 ÀF)
IP65 with plug complying with
EN 175301-803-A
IP67 with M12 plug
IP67 with cable
IP67 with cable fast-fit gland
complying with EN 61326-1
complying with EN 61326-2-3
complying with NAMUR NE21,
only for ATEX device and max.
measured value deviation of
< 1 %
approx. 0.090 kg (0.198 lb)
Dimension drawings
Plug complying with EN 175301-
803-A
Form A with cable inlet
M16x1.5 or
-14NPT or Pg 11
M12 plug
2- or 3-wire (0.5 mm
2
)
Cable (± 5.4 mm)
Fast-fit cable gland
Media temperature
-15 ... +125 ÀC (+5 ⁄ +257 ÀF)
-35 ... +100 ÀC < 100 bar
(-31 ⁄ +212 ÀF; < 1 450 psi)
-20 ... +100 ÀC (-4 ⁄ +212 ÀF)
-40 ... +145 ÀC < 100 bar
(-40 ⁄ +293 ÀF; < 1 450 psi),
can be used for drinking water
SITRANS P210 (7MF1566)
A5E03304388-3, 05/2012

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

7mf1566

Inhaltsverzeichnis