Metrohm 765 Dosimat Gebrauchsanweisung Seite 45

Inhaltsverzeichnis

Werbung

The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
4.2.7).
4.2.30 VDL VALUE
Value = 0.001 ... 999.999
Setting the diluting volume in ml.
In mode DIL only, in other modes the command is not accepted and bit 0 of information
byte 2 is set to 1.
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
4.2.7).
4.2.31 VLI VALUE
Value = 0.001 ... 999.999
Setting the security volume in ml.
In modes DOS, DIS C, and PULSE only, in other modes the command is not accepted
and bit 0 of information byte 2 is set to 1.
The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the
volume of the cylinder:
765 Dosimat
Cylinder
Volume
ml
ml/min
1
0.001
5
0.001
10
0.001
20
0.002
50
0.005
or volume
resp. and sets bit 1 of information byte 2 to 1 (see
max
min,
Diluting volume, not live
Cylinder
Volume
ml
ml/min
1
0.001
5
0.001
10
0.001
20
0.002
50
0.005
or volume
resp. and sets bit 1 of information byte 2 to 1 (see
max
min,
Security volume, not live
4.2 Control commands
Volume
min
max
ml/min
0.900
4.9009
9.800
19.700
49.500
Volume
min
max
ml/min
999.999
999.999
999.999
999.999
999.999
41