Trouble Shooting
Trouble
Broken pipette in the
adapter.
The instrument does not
not draw in liquid/pipette
drips.
Liquid in the adapter and
the suction tube.
Pipette not held properly
in the adapter.
Possible Cause
The pipette was damaged/
forced into the adapter.
Adapter or suction tube
damaged.
The thumb wheel was
turned too quickly.
Adapter is wet or greasy.
– 18 –
Action to be taken
Remove the glass splinters by means of
forceps, replace adapter and suction tube,
if necessary, ➤ page 16.
Replace adapter and suction tube,
➤ page 16.
Clean or disinfect adapter and/or suction tube,
if necessary, replace,
➤ page 14.
Clean or disinfect adapter and suction tube, and
allow to dry completely. If necessary, replace,
➤ page 14.