Herunterladen Inhalt Inhalt Diese Seite drucken

PR 9116 Handbuch Seite 60

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

9116 Universal converter
CONNEC
Selected sensor connection type for RTD or linear resistance
measurements (for SENSOR = Ni or Pt or IN.TYPE = LIN.R):
2W = 2-wire
3W = 3-wire
4W = 4-wire
If 2W or 3W is selected, the end user must ensure that the
applied sensor wiring does not introduce failures exceeding the
requirements for the safety application.
R 0%
0% input range for linear resistance measurements
(for IN.TYPE = LIN.R).
This value must be < (R 100% - minimum span), refer to 4.2.1
R 100%
100% input range for linear resistance measurements
(for IN.TYPE = LIN.R)
This value must be > (R 0% + minimum span), refer to 4.2.1
UNIT
Selected temperature unit for IN.TYPE = TEMP
°C = degrees Celsius
°F = degrees Fahrenheit
For IN.TYPE ≠ TEMP selectable units refer to routing diagram
DISP.LO
0% display value on 4501. Can be used as base for relay
setpoints (for IN.TYPE ≠ TEMP)
DISP.HI
100% display value on 4501. Can be used as base for relay
setpoints (for IN.TYPE ≠ TEMP)
NEW.PAS
Password for protection of the device configuration from
unauthorized access. Range from 0 to 9999.
Version No. V3R0
safety Manual