Herunterladen Inhalt Inhalt Diese Seite drucken

LumaSense Technologies IMPAC IP 140-LO Betriebsanleitung Seite 25

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

IMPAC pyrometer IP 140-LO
Baud rate:
Temp. display °C / °F
Wait time:
Internal temperature:
Max. internal
temperature:
Error status:
Laser targeting light:
Reading interface:
Lock keyboard:
Reading parameters:
Device type:
Serial number:
Device type /
software version:
Detailed Software
version:
Reference number:
Note: the letter "l" means the lower case letter of "L".
Additional instruction for the RS 485 interface:
Requirements to the master system during half-duplex operation:
1. After an inquiry, the bus should be switched into a transmission time of 3 bits
(some older interfaces are not fast enough for this).
2. The pyrometer's response will follow after 5 ms at latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated.
AAbrX
X=1...6 or 8 (dec.)
AAfhX
Output: X = 0: display in °C; X = 1: display in °F
AAtwXX
XX = 00 ... 99 (decimal, in bit time of the current baud rate)
AAgt
Output: XXX (dec. 000 ... 099°C, 032 ... 210°F
AAtm
Output: XX (dec. 00 ... 99, always in °C)
AAfs
Output 1 byte hex (00 = no error)
AAlaX
X = 0 switch off laser
X = 1 switch on laser
AAin
Output: 1 or 2 (1 = RS232, 2 = RS485)
)
AAlkX *
X = 0 ... 3
AApa
Output 11-digit, decimal:
AAna
Output: "IP 140" (16 ASCII-characters)
AAsn
Output: XXXXX (dec. 5-digit)
AAve
Output: XXYYZZ (6-digit decimal)
AAvs
tt.mm.yy XX.YY
tt = Day; mm = month; yy = year; XX.YY = software version
AAbn
Output: XXXXXX (hex 6-digit)
1 = 2400 baud
5 = 38400 baud
2 = 4800 baud
6 = 57600 baud
3 = 9600 baud
(7 is not available)
4 = 19200 baud
8 = 115200 baud
Bit 0 = 1: Measurement unit doesn't work
Bit 1 = 1: Internal temperature measurement
doesn't work
1 = lock lk1, removal with command lk0 or power off-on
0 = removal of lock lk1
3 = continuous lock lk3, removal only with command lk2
2 = removal of lock lk3
Digit 1 und 2 (10...99 or 00): Emissivity
Digit 3 (0 ... 6): Exposure time
Digit 4 (0 ... 8): Clear time max. storage
Digit 5 (0 ... 1): Analog output
Digit 6 und 7: (00 ... 98): Internal temperature
Digit 8 und 9 (00 ... 97): Address
Digit 10 (0 ... 6 or 8): Baud rate
Digit 11 (0, 1, 3): Keyboard lock
XX = 63 (IP 140)
YY = Month of software version
ZZ = Year of software version