6.2.2 Align pyrometer

The pyrometer is equipped with an LED pilot light for accurate alignment of the sensor with the object to be
measured. Align the pyrometer with the object to be measured as follows:
1. Switch on the supply voltage to the pyrometer.
2. Allow a starting time of approx. 5 minutes for thermo-stabilization. Than the pyrometer has stabilized and
is ready to work with the given accuracy.
3. Switch on the pilot light by pressing the push button at the rear side. Make sure that there is nothing in
the path of rays.
Note:
In order to avoid measuring errors, the area of the pilot light must not be larger than the object to be
measured.
4. If you have a pyrometer with vario optic head, adjust the necessary measuring distance (see Setting the
optic (section 7.1) at page 16).
Note:
The fixed optics cannot be adjusted.
The pyrometer is thus aligned and ready for temperature measurement.