Table 4 Warnings
Warning
Description
Laser current
Occurs when the laser current exceeds ±30% of the nominal
level.
Buffer full
The instrument is powered on and the buffer is full. Old data
will be overwritten.
Flow system
A Flow Alarm warning occurs when the firmware is unable to
maintain a steady state.
Overconcentration Occurs when concentration limits of the sensor are
detected. The warning clears when concentration values
return to normal.
Calibration failure
Occurs when the calibration signal registers a sensitivity
change of ± 10% size error in the first channel.