320 XT/XB/IP65
3.7.5 Assembling the support piece, the inner cone and the weight holders
1. Introduce the support piece with the loosely held coupling anchor over the two inner assem-
bly pins [9].
2. It must be possible to push the support piece smoothly over the two inner assembly pins. In
case the support piece can not be pushed smoothly running over the two assembly pins, uns-
crew the pin [9] and tighten it again.
3. Locate the support piece and the balance arm using the coupling anchor [13] and the two
screws [14] → The support piece and balance arm are fixed now.
4. Assemble the inner cone (12/p.12) to the support piece with the 3 screws (23/p.12).
SCS only:
5.
Assemble the two weight holders (9/p.12) and (10/p.12) .
Precisa Gravimetrics AG
[13]
[9]
[14]
Index d1
Twin Reference Weight Version
[9]
47