Section 11: Specifications
Cavitron™ PROPHY-JET ® Air Polishing
Prophylaxis System: 100/115 VAC Systems
Mode of Operation: Continuous operation with intermittent loading:
10 minutes ON / 20 minutes OFF
Electrical Voltage 100V Gen 122J
115V Gen 122
Current
1.1/1.1 amps @ 100v/115v
Frequency
50/60 Hertz
Water Pressure
25–60 psig
Air Pressure
65–100 psig (448–690 kPa)
Flow Rate
Minimum setting (CCW) 0–15 ml/min
Maximum setting (CW) >55 ml/min
Weight
12 lb. (5,4 kg)
Dimensions
Height:
Width:
Depth:
Handpiece Cable:
Foot Control Cable:
Power Cord:
Water Supply Line:
Air Supply Line:
®
Cavitron™ PROPHY-JET
Prophylaxis System: 230 VAC Systems
Mode of Operation: Continuous operation with intermittent loading:
10 minutes ON / 20 minutes OFF
Electrical Voltage 230V Gen 122U
Current
.7/.6 amps @ 230v
Frequency
50/60 Hertz
Water Pressure
172–414 kPa
Air Pressure
448–690 kPa
Flow Rate
Minimum setting (CCW) 0–15 ml/min
Maximum setting (CW) >55 ml/min
Weight
5,4 kg
Dimensions
Height:
Width:
Depth:
Handpiece Cable:
Foot Control Cable:
Power Cord:
Water Supply Line:
Air Supply Line:
Section 12: Classification
• Type of protection against electric shock:
• Degree of protection against electric shock:
• Degree of protection against the harmfulingress of water:
• Mode of operation:
• Degree of safety of application in the presence of a flammable
anaesthetic mixture with air or with oxygen or nitrous oxide:
• According to medical device directive:
Section 13: Disposal of Unit
• Accordance with local and state laws.
ENGLISH
14
(172–414 kPa)
4 in.
7 3/8 in.
9 1/2 in.
81 in.
96 in.
84 in.
98 in.
Air Polishing
10,2 cm
18,7 cm
24,1 cm
2,0 m
2,4 m
2,1 m
2,5 m
OPERATING ENVIRONMENT
Temperature
Relative Humidity
TRANSPORT AND STORAGE CONDITIONS
Temperature
Relative Humidity
Atmospheric Pressure
96 in.
2,4 m
Class 1
Type B
Ordinary
Continuous
Equipment not suitable for use in the presence
of flammable anesthetics or oxygen.
IIA (rule 9)
10 to 25 degrees celsius
30 to 75 percent ( non-condensing )
0 to 70 degrees celsius
10 to 95 percent ( non-condensing )
500 HPa to 1060 Hpa
AC POWER
IPX1 classification of increase of water
TYPE B EQUIPMENT
FOOTSWITCH
Footswitch not for operating theatres
Protection Class- IPX1
Refer to instruction manual
for information
POWER ON
POWER OFF