After repro-
cessing
Remove the silicone hose once reprocessing is complete. Then re-
move the transmission instrument from the A 803 or AUF 1 holder.
After processing dry internal sections of transmission instruments
with sterilised compressed air and then sterilise as required follow-
ing manufacturer's instructions and national health and safety regu-
lations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.