Herunterladen Inhalt Inhalt Diese Seite drucken

IKA IKAMAG RET control-visc C Betriebsanleitung Seite 38

Inhaltsverzeichnis

Werbung

X = 8
Pt temperature sensor – safety temperature
X = 50 Pt100 temperature sensor offset in K
X = 52 Pt1000 1st temperature sensor offset in K
X = 53 Pt1000.2nd temperature sensor offset in K
NAMUR instructions
IN_NAME
IN_PV_X
X=1;2;3;4;5;
7; 8;
IN_SOFTWARE
IN_SP_X
X=1;2;3;4;7;8;
12;42;50;52;53;
54;55;56;
IN_TYPE
OUT_NAME name
OUT_SP_12@n
OUT_SP_42@n
OUT_SP_X n
X=1;2;4;7;50;
52;53;54;55;56
OUT_WD1@m
OUT_WD2@m
RESET
38
(-3.0 <= n <=+3.0 )
(-3.0 <= n <=+3.0 )
(-3.0 <= n <=+3.0 )
Function
Input description name
Reading the real value
Input software ID number, date and version
Reading the set rated value
Input laboratory unit ID
Output description name. (Max. 6 characters, default: IKAMAG)
Setting the WD safety temperature with the echo of the set value
Setting the WD safety speed with the echo of the set value
Setting the rated value to n
Watchdog mode 1: When a WD1 event occurs, the heating and stirring functions are shutdown and message ER 2 is
displayed.
Set the watchdog time to m (20...1500) seconds, with echo of the watchdog time. This instruction starts the watchdog
function and must be sent within the set watchdog time.
Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD safety setpoint speed and the
temperature setpoint will be set to the WD safety setpoint temperature. The WD warning is displayed.
The WD2 event can be reset with OUT_WD2@0 - resetting also blocks the watchdog function. Set the watchdog time
to m (20...1500) secondes, with echo of the watchdog time. This command starts the watchdog function and must be
sent within the set watchdog time.
Switching off the instrument function.
X = 54 Error 5 response time in seconds
(180 <= n <= 1200)
X = 55 Intermittent mode cycle time in seconds
(10 <= n <= 600)
X = 56 Intermittent mode, duration of interruption in
seconds
( 5 <= n <= 60)
Display
(additional)
RETCV_0107

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ikamag ret control-viscIkamag ret control-visc c

Inhaltsverzeichnis