9.4

Settings

The following settings are programmed at the factory:
SetBasics
(Basic settings)
Decimalpoint
Display Refresh
Measurement Unit
Back to Menu
SetFunction
Analog Signal
Output Pulse
Set min Q
Set max Q
Totalcount Reset
Back to Menu
SetCalibration
(Setting of calibration values)
K-Factor
ZeroOffset
Adjust 4 mA
Adjust 20 mA
Back to Menu
Service
Version
Sensor Check
Service A
Status
Status
3021
Definition
Setting of number of decimal points
Repeat time of display refreshing
Setting of measurement unit
Back to main menu
Definition
Selection of current output
Definition of frequency output
Input of minimum flow corresponding
to 0 / 4 mA
Input of maximum flow corresponding
to 20 mA
Totalizer can be reset
Back to main menu
Definition
Input of pulses / litre - pulse rate
Offset conditional on turbine - zero
point correction
Adjustment of current output to 4 mA
- change only for recalibration!
Adjustment of current output to
20 mA - change only for recalibration!
Back to main menu
Definition
Indication of software version
Check of sensor
Service number is displayed
Programming mode can be locked
Input of code number
36 / 52
Values / Unit / Remark
0000.xx (= two decimal points)
0.5 s
L / h
Values / Unit / Remark
4 - 20 mA
Pulse / L: 1 pulse / L or
L / Pulse: 1 L / pulse
0 L / h
3600 L / h (DN 25) - 25000 L / h
(DN 50)
Values / Unit / Remark
Set at the factory, see test certificate
Set at the factory, see test certificate
Set at the factory
Set at the factory
Values / Unit / Remark
Carried out at the factory
unlocked
Code: 0