Widerstandsthermometer LTS 190
MANUAL
Resistance thermometer LTS 190
5.2.2 Fine adjustment by correction potentiometer
As can be seen in the picture above, the correction potentiometer is located on the top of
the KMUS. This can be used for fine adjustment of the output voltage. A seal secures the
potentiometer against accidental adjustment.
6. Technical specifications
Material protective fitting:
Length of sensor:
Connecting box:
Dimensions:
Protection class:
Operating temperature:
6.1 Technical data DMU50
Operating temperature:
Operating voltage:
Electricity demand:
Input:
Measuring range max.:
Measuring span min.:
Accuracy:
Output:
Sensor break:
Standard configuration:
Max. permissible load:
Display:
Alignment display:
Display-digits:
Displayrange:
Electrical connection:
Configuration:
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 •
stainless steel 1.4571 (V4A)
45 mm other lengths on request
plastic polyamide
58 x 64 x 34 mm
IP 65 according to DIN 60529
-50 °C bis +130 °C
(deviation when using a transmitter)
-30 °C bis +70 °C
UB = 10..35 V DC
7,3 mA (UB=24V) + 4..20mA output
PT1000 2-wire
-100°C bis +650°C
10K
<+-0,1% from final value
4..20mA 3-wire active (underflow 3,5mA,overflow
20,5mA)
21mA
4mA = -50°C, 20mA = 70°C
(more temperature ranges parameterizable by factory)
Rmax=[(UB – 6V) / 0,021 A] Ω
high-resolution OLED display 0,96 Zoll
0° oder 180°
4-digits
-99,9 bis +999,9°C
5x terminal connection 1,5 mm²
factory-side