4 Pipetting
Discharge liquid (pipettes calibrated to contain "Ex")
b.
a. Hold the instrument upright, and place the pipette tip against the wall
of a vessel (inclined position of the vessel).
b. Press the venting button until the liquid has run out.
c. As soon as the meniscus comes to a standstill in the pipette tip, the
waiting time (if applicable) starts.
d. Then take the pipette out of the vessel by wiping it against the wall of
the vessel. During this procedure the residue of liquid partly runs out.
Removing the pipette
Do not touch the contaminated area of the pipette. Wear
protective gloves.