The 6082 MX-PRO R3 ambulance cot is designed to be power washable. The unit may show some signs
of oxidation or discoloration from continuous washing, however, no degradation of the cots performance char‐
acteristics or functionality will occur due to power washing as long as the proper procedures are followed.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate.
(Appropriate disinfectant for nylon Velcro should be determined by the hospital.)
Washing Limitations
WARNING
Use any appropriate personal safety equipment (goggles, respirator, etc.) to avoid the risk of inhaling conta‐
gion. Use of power washing equipment can aerate contamination collected during the use of the cot.
CAUTION
Failure to properly clean or dispose of contaminated mattress or cot components will increase the risk of expo‐
sure to bloodborne pathogens and may cause injury to the patient or the operator.
S
DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE UNI
S
Maximum water temperature should not exceed 180 degrees F/ 82 degrees C.
S
Maximum air dry temperature (cart washers) is 240 degrees F /115 degrees C.
S
Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand held wand is being used
to wash the unit, the pressure nozzle must be kept a minimum of 24 inches/61 cm from the unit.
S
Towel dry all casters and interface points.
S
After washing the cot, follow the lubrications procedure on the following page.
S
Failure to comply with these instructions may invalidate any/all warranties.